Voltage dependent pulse shape analysis of Geiger-Müller counter

B. Almutairi, T. Akyurek, S. Usman*

*Bu çalışma için yazışmadan sorumlu yazar

Araştırma sonucu: ???type-name???Makalebilirkişi

11 Atıf (Scopus)

Özet

Detailed pulse shape analysis of a Geiger-Müller counter is performed to understand the pulse shape dependence on operating voltage. New data is presented to demonstrate that not all pulses generated in a GM counter are identical. In fact, there is a strong correlation between the operating voltage and the pulse shape. Similar to detector deadtime, pulse shapes fall in three distinct regions. For low voltage region, where deadtime was reported to reduce with increasing voltage, pulse generated in this region was observed to have a fixed pulse width with a variable tail. The pulse width and fall time of the tail was observed to be a function of applied voltage; exponentially reducing with increasing voltage with an exponent of negative 6E-04 and 2E-03 respectively. The second region showed a pulse without any significant tail. During this time the detector deadtime was earlier reported to be at its minimum. The highest voltage region demonstrated a different deadtime mechanism where the second pulse was reduced in width. During this time the deadtime seemed to be increasing with increasing voltage. This data allows us to gain some unique insight into the phenomenon of GM detector deadtime not reported thus far.

Orijinal dilİngilizce
Sayfa (başlangıç-bitiş)1081-1090
Sayfa sayısı10
DergiNuclear Engineering and Technology
Hacim51
Basın numarası4
DOI'lar
Yayın durumuYayınlandı - Tem 2019
Harici olarak yayınlandıEvet

Bibliyografik not

Publisher Copyright:
© 2019

Parmak izi

Voltage dependent pulse shape analysis of Geiger-Müller counter' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.

Alıntı Yap