Verification of LDR brachytherapy dose distributions for different source geometries by Monte Carlo method

Tercüme edilen katkı başlığı: İnterstisyel LDR brakiterapide farklı kaynak geometrileri için doz daǧılımlarının monte carlo yöntemiyle doǧrulanması

Hediye Acun*, Ahmet Bozkurt, Gönül Kemikler, Ismail Özbay, Işik Aslay

*Bu çalışma için yazışmadan sorumlu yazar

Araştırma sonucu: Dergiye katkıMakalebilirkişi

Özet

OBJECTIVES: This study aimed to verify dose distributions for 192Ir wires obtained from both the CMS XiO treatment planning system (TPS) and the Monte Carlo (MC) method. METHODS: Using 192Ir source of 1.84 mCi/cm activity, three different wire setups were investigated. Whole properties of the wires were modelled in the MC code MCNP. Dose distributions and isodose curves were obtained and then compared with TPS results. RESULTS: Isodose curves from TPS and MC were found to differ as 0.28 mm, 0.4 mm and 0.57 mm for one, two and three wire applications, respectively. MC was observed to yield much narrower isodose curves. CONCLUSION: The greater the number of wires, the larger the differences between the isodose widths calculated with TPS and MC. However, these differences are less than 1 mm. The results show that, for low dose rate (LDR) brachytherapy applications, CMS XiO TPS and MC yield consistent isodose curves.

Tercüme edilen katkı başlığıİnterstisyel LDR brakiterapide farklı kaynak geometrileri için doz daǧılımlarının monte carlo yöntemiyle doǧrulanması
Orijinal dilİngilizce
Sayfa (başlangıç-bitiş)181-188
Sayfa sayısı8
DergiTurk Onkoloji Dergisi
Hacim27
Basın numarası4
DOI'lar
Yayın durumuYayınlandı - 2012
Harici olarak yayınlandıEvet

Parmak izi

İnterstisyel LDR brakiterapide farklı kaynak geometrileri için doz daǧılımlarının monte carlo yöntemiyle doǧrulanması' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.

Alıntı Yap