Ana gezinime geç Aramaya geç Ana içeriğe geç

Three-Dimensional Finite Element Analysis of a Notched Insert Design for Reverse Total Shoulder Arthroplasty to Prevent Scapular Notching

  • Nezih Ziroglu*
  • , Hüseyin Balin
  • , Vahdet Ucan
  • , Ergun Bozdag
  • , Mehmet Kapicioglu
  • , Kerem Bilsel
  • *Bu çalışma için yazışmadan sorumlu yazar
  • Acibadem Mehmet Ali Aydinlar Universitesi
  • Istanbul Technical University
  • Bezmialem Vakif University

Araştırma sonucu: Dergiye katkıMakalebilirkişi

Özet

Purpose: Reverse total shoulder arthroplasty (RSA) is an effective treatment option for rotator cuff arthropathy. Scapular notching following RSA remains a major complication and has a high incidence. This finite element analysis (FEA) study provides a future reference for the optimal design of the insert component of RSA. This study aims to clarify the effect of a new design RSA with a notched insert on the range of adduction, scapular notching, and stress variation of its insert component using three-dimensional (3D) FEA. Methods: 3D nominal Grammont-type monobloc RSA implant components are modeled on the sawbones glenohumeral joint. The polyethylene insert is redesigned with notching of the inferior part. The comparison of standard and notched designs was performed by FEA for stress pressure of scapular notching and the degree of adduction. 3D mesh models are created for stress analysis to compare the results between standard and notched inserts for the adduction. Results: The redesigned notched inserts had an additional ~ 11.2° on adduction and prevented scapular notching. The stress analysis results for the notched insert design were lower than the standard ones (4.7 vs 22.4 Kpa). Conclusions: Notched insert design of Grammont-type RSA could provide additional adduction with lower stress on the glenoid, leading to less scapular notching. Further experimental and clinical studies on different RSA types are needed to verify this effect. Study Design: Basic Science Study; Biomechanics and Computer Modeling.

Orijinal dilİngilizce
Sayfa (başlangıç-bitiş)1874-1880
Sayfa sayısı7
DergiIndian Journal of Orthopaedics
Hacim57
Basın numarası11
DOI'lar
Yayın durumuYayınlandı - Kas 2023

Bibliyografik not

Publisher Copyright:
© 2023, Indian Orthopaedics Association.

Parmak izi

Three-Dimensional Finite Element Analysis of a Notched Insert Design for Reverse Total Shoulder Arthroplasty to Prevent Scapular Notching' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.

Alıntı Yap