Özet
In this study, in order to eliminate the drawbacks of elution step and to reach higher enrichment factors, a novel preconcentration/separation technique for the slurry analysis of sorbent loaded with lead prior to its determination by electrothermal atomic absorption spectrometry was described. For this purpose, at first, lead was collected on ethylene glycol dimethacrylate methacrylic acid copolymer (EGDMA-MA) treated with ammonium pyrolidine dithiocarbamate (APDC) by conventional batch technique. After separation of liquid phase, slurry of the sorbent was prepared and directly pipetted into graphite furnace of atomic absorption spectrophotometer. Optimum conditions for quantitative sorption and preparation of the slurry were investigated. A 100-fold enrichment factor could be easily reached. The analyte element in certified sea-water and Bovine-liver samples was determined in the range of 95% confidence level. The proposed technique was fast and simple and the risks of contamination and analyte loss were low. Detection limit (3δ) for Pb was 1.67 μg l-1.
| Orijinal dil | İngilizce |
|---|---|
| Sayfa (başlangıç-bitiş) | 804-808 |
| Sayfa sayısı | 5 |
| Dergi | Journal of Hazardous Materials |
| Hacim | 150 |
| Basın numarası | 3 |
| DOI'lar | |
| Yayın durumu | Yayınlandı - 11 Şub 2008 |
BM SKH
Bu sonuç, aşağıdaki Sürdürülebilir Kalkınma Hedefine/Hedeflerine katkıda bulunur
-
SKH 14 Sudaki Yaşam
Parmak izi
Slurry analysis after lead collection on a sorbent and its determination by electrothermal atomic absorption spectrometry' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.Alıntı Yap