Portable spectroscopic fast neutron probe and 3He detector dead-time measurements

T. Akyurek*, L. P. Tucker, X. Liu, S. Usman

*Bu çalışma için yazışmadan sorumlu yazar

Araştırma sonucu: ???type-name???Makalebilirkişi

9 Atıf (Scopus)

Özet

This paper presents dead-time calculations for the Portable Spectroscopic Fast Neutron Probe (N-Probe) using a combination of the attenuation law, MCNP (Monte Carlo N-particle Code) simulations and the assumption of ideal paralyzing and non-paralyzing dead-time models. The N-Probe contains an NE-213 liquid scintillator detector and a spherical 3He detector. For the fast neutron probe, non-paralyzing dead-time values were higher than paralyzing dead-time values, as expected. Paralyzing dead-time was calculated to be 37.6 μs and non-paralyzing dead-time was calculated to be 43.7 μs for the N-Probe liquid scintillator detector. Dead-time value for Canberra 3He neutron detector (0.5NH1/1K) was also estimated using a combination of subcritical assembly measurements and MCNP simulations. The paralyzing dead-time was estimated to be 14.5 μs, and the non-paralyzing dead-time was estimated to be 16.4 μs for 3He gas filled detector. These results are consistent with the dead-time values reported for helium detectors.

Orijinal dilİngilizce
Sayfa (başlangıç-bitiş)15-21
Sayfa sayısı7
DergiProgress in Nuclear Energy
Hacim92