Efficiency calibration of a well-type Ge detector for voluminous samples in cylindrical geometry

Ayse Nur Solmaz, Dogan Bor, Haluk Yücel

Araştırma sonucu: Kitap/Rapor/Konferans Bildirisinde BölümKonferans katkısıbilirkişi

Özet

To study environmental samples, a 44.8% relative efficient well-type Ge detector with an active volume of 218 cm3 was calibrated in three different voluminous sample geometries. The models were tested by absolute full energy peak efficiency calibration with use of certified multinuclide standards. The efficiency results corrected for this experiment obtained self-absorption factors from the simple photon transmission experiments by using point-gamma sources for small cylindrical geometry and theoretically calculated self absorption factors based on the database related to mass attenuation coefficients for vial geometry. Tufa, soil, and marble were also measured at these counting geometries. The activity results obtained with these models were compared with each other.

Orijinal dilİngilizce
Ana bilgisayar yayını başlığıNew Techniques for the Detection of Nuclear and Radioactive Agents
YayınlayanSpringer Verlag
Sayfalar193-205
Sayfa sayısı13
ISBN (Basılı)9781402095986
DOI'lar
Yayın durumuYayınlandı - 2009
Harici olarak yayınlandıEvet

Yayın serisi

AdıNATO Science for Peace and Security Series B: Physics and Biophysics
ISSN (Basılı)1874-6500