Characterization of plasma sprayed bioglass coatings on titanium

Gultekin Goller*, Faik N. Oktar, Tokay Yazici, Sebahattin Gurmen, Eyup Sabri Kayali

*Bu çalışma için yazışmadan sorumlu yazar

Araştırma sonucu: ???type-name???Konferans makalesibilirkişi

9 Atıf (Scopus)

Özet

Bioglass, 45S5 containing 45 % SiO2, 6 % P2O5, 24.5 % CaO and 24.5 % Na2O all in weight percent, was plasma sprayed onto a titanium substrate with and without 60 Al2O3 40 TiO2 as a bond coating. Mechanical properties were evaluated by following the ASTM C633 method and the microstructural characterization has been carried out by using scanning electron microscope. Results indicate that the bonding strength of bond coatings is three times higher than of the coatings that are directly applied to the surface. It has been observed that there is a uniform coating layer with a thickness of 110 μm and that there was not any reaction observed at the coating-metal interface.

Orijinal dilİngilizce
Sayfa (başlangıç-bitiş)283-286
Sayfa sayısı4
DergiKey Engineering Materials
Hacim240-242
DOI'lar
Yayın durumuYayınlandı - 2003
EtkinlikProceedings of the 15th International Symposium on Ceramics in Medicine; The Annual Meeting of the International Society for Ceramics in Medicine - Sydney, NSW, Australia
Süre: 4 Ara 20028 Ara 2002

Parmak izi

Characterization of plasma sprayed bioglass coatings on titanium' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.

Alıntı Yap