Characterization of Open-Ended Coaxial Probe Sensing Depth with Respect to Aperture Size for Dielectric Property Measurement of Heterogeneous Tissues

Cemanur Aydinalp, Sulayman Joof, Ismail Dilman, Ibrahim Akduman, Tuba Yilmaz*

*Bu çalışma için yazışmadan sorumlu yazar

Araştırma sonucu: ???type-name???Makalebilirkişi

17 Atıf (Scopus)

Özet

The open-ended coaxial probe (OECP) method is frequently used for the microwave dielectric property (DP) characterization of high permittivity and conductivity materials due to inherent advantages including minimal sample preparation requirements and broadband measurement capabilities. However, the OECP method is known to suffer from high measurement error. One well-known contributor to the high error rates is tissue heterogeneity, which can potentially be managed through the selection of a probe with a proper sensing depth (SD). The SD of the OECP is dependent on many factors including sample DPs and probe aperture diameter. Although the effects of sample DPs on SD have been investigated to some extent in the literature, the probe aperture diameters, particularly small diameters, have not been fully explored. To this end, the SDs of probes with three different apertures (0.5, 0.9 and 2.2 mm-diameters) were analyzed in this study. Probes’ SDs were first investigated with simulations using a double-layered sample configuration (skin tissue and olive oil). Next, experiments were performed using a commercial OECP with a 2.2 mm aperture diameter. The SD was categorized based on 5%, 20% and 80% DP change. Among these threshold values, a 5% DP change was selected as the benchmark for SD categorization. The findings suggest that probes with a smaller aperture size and correspondingly smaller SD should be utilized when measuring the DPs of thin and multilayered samples, such as healthy and diseased skin tissues, to increase the measurement accuracy.

Orijinal dilİngilizce
Makale numarası760
DergiSensors
Hacim22
Basın numarası3
DOI'lar
Yayın durumuYayınlandı - 1 Şub 2022

Bibliyografik not

Publisher Copyright:
© 2022 by the authors. Licensee MDPI, Basel, Switzerland.

Finansman

Funding: This work has received funding from Scientific and Technological Research Council of Turkey under grant agreement 118S074 and COST Action grant agreement CA17115.

FinansörlerFinansör numarası
European Cooperation in Science and TechnologyCA17115
Türkiye Bilimsel ve Teknolojik Araştirma Kurumu118S074

    Parmak izi

    Characterization of Open-Ended Coaxial Probe Sensing Depth with Respect to Aperture Size for Dielectric Property Measurement of Heterogeneous Tissues' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.

    Alıntı Yap