Özet
This study presents a new dead-time measurement method using the gamma attenuation law and generalized dead-time models for nuclear gamma-ray detectors. The dead-time of the NaI(Tl) detection system was obtained to validate the new dead-time determination method using very thin lead and polyethylene absorbers. Non-paralyzing dead-time was found to be 8.39 μs, and paralyzing dead-time was found to be 8.35 μs using lead absorber for NaI(Tl) scintillator detection system. These dead-time values are consistent with the previously reported dead-time values for scintillator detection systems. The gamma build-up factor's contribution to the dead-time was neglected because a very thin material was used.
Orijinal dil | İngilizce |
---|---|
Sayfa (başlangıç-bitiş) | 4093-4097 |
Sayfa sayısı | 5 |
Dergi | Nuclear Engineering and Technology |
Hacim | 53 |
Basın numarası | 12 |
DOI'lar | |
Yayın durumu | Yayınlandı - Ara 2021 |
Harici olarak yayınlandı | Evet |
Bibliyografik not
Publisher Copyright:© 2021 Korean Nuclear Society