Abstract
In this study, we investigated the validity of using a two-dimensional gait analysis system, which employed industrial cameras and a custom-developed Inertial Measurement Unit system, compared to the gold standard three-dimensional Vicon gait analysis system. Four subjects with a mean age of 26.76 (SD: 0.71) were included in the study, and gait data was collected from their 8 limbs. Correlation analyses examined the relationship between the data obtained from the two-dimensional gait analysis system and the three-dimensional Vicon system. The study's results showed a high correlation for the hip and knee joints between the two systems, but a lower correlation for the ankle joint. This pilot study demonstrated the potential of using industrial cameras and IMU systems for practical gait analysis.
| Original language | English |
|---|---|
| Title of host publication | TIPTEKNO 2024 - Medical Technologies Congress, Proceedings |
| Publisher | Institute of Electrical and Electronics Engineers Inc. |
| ISBN (Electronic) | 9798331529819 |
| DOIs | |
| Publication status | Published - 2024 |
| Event | 2024 Medical Technologies Congress, TIPTEKNO 2024 - Mugla, Turkey Duration: 10 Oct 2024 → 12 Oct 2024 |
Publication series
| Name | TIPTEKNO 2024 - Medical Technologies Congress, Proceedings |
|---|
Conference
| Conference | 2024 Medical Technologies Congress, TIPTEKNO 2024 |
|---|---|
| Country/Territory | Turkey |
| City | Mugla |
| Period | 10/10/24 → 12/10/24 |
Bibliographical note
Publisher Copyright:© 2024 IEEE.
Keywords
- Industrial cameras
- Inertial Measurement Unit system
- Two-dimensional gait analysis system