Abstract
A non-ionic micellar electrokinetic chromatography method coupled to laser-induced fluorescence detection was developed for the quantitative determination of biogenic amines. Complete resolution of six biogenic amines - FITC derivatives - was achieved in less than 10 min, employing 10mM Brij 35 in 75mM borate buffer and at pH 9.7 as the running electrolyte. Relative fluorescence intensities of biogenic amines enhanched considerably and separation time decreased considerably when Brij 35 was substituted for SDS in the same buffer. The detection limits of the method for biogenic amines were found between 0.416 and 1.26 nM. The precisions for the corrected peak areas were calculated as between 1.63 and 3.24 %RSD. The applicability of the method was demonstrated by analyzing biogenic amines in the processed, brined, and dry-salted fish samples. The method is simple and rapid, and widely applicable for the determination of biogenic amines in food samples.
Original language | English |
---|---|
Pages (from-to) | 2174-2179 |
Number of pages | 6 |
Journal | Electrophoresis |
Volume | 31 |
Issue number | 13 |
DOIs | |
Publication status | Published - Jul 2010 |
Keywords
- Biogenic amines
- Brij 35
- Fish
- Food
- MEKC