Design of Current and Transimpedance-Mode Instrumentation Amplifier for Biomedical Applications Using a Single Current Conveyor Transconductance Amplifier

  • Emre Ozer*
  • , Abdullah Yesil
  • , Firat Kacar
  • *Corresponding author for this work
  • Bandırma Onyedi Eylül University
  • Istanbul University - Cerrahpaşa

Research output: Contribution to journalArticlepeer-review

Abstract

This paper presents a new instrumentation amplifier employing a single current conveyor transconductance amplifier (CCTA) and two PMOS transistors. The proposed circuit operates in both current-mode and transimpedance-mode. Electronic adjustment feature is provided by using an electronic resistor realized with two transistors. The SPICE simulation results based on 0.13-μm IBM CMOS parameters are carried out to confirm the theoretical analysis. The proposed IA operates at ±0.9 V supply voltage and the total power consumption is found to be 0.9 mW. The functionality and robustness of the proposed circuit has been verified by frequency and time-domain simulations. As an application example, the proposed current-mode instrumentation amplifier has been employed for removing the noise component in electrocardiogram (ECG) signals. Moreover, the proposed circuit underwent experimental testing utilizing commercially accessible integrated circuits (ICs), namely the AD844 and LM13700.

Original languageEnglish
JournalInternational Journal of Circuit Theory and Applications
DOIs
Publication statusAccepted/In press - 2026

Bibliographical note

Publisher Copyright:
© 2026 John Wiley & Sons Ltd.

Keywords

  • CCTA
  • CMRR
  • current-mode
  • ECG application example
  • instrumentation amplifier
  • transimpedance-mode