Analog filters for Enhanced Signal Reception of Magnetic Particle Imaging (MPI) Scanner

M. Irfan, H. Mossa, N. Dogan

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

This paper presents the analog filters for the receive chain of the magnetic particle imaging (MPI) scanner. Direct feedthrough interference between drive and receive coil was evaluated with three analog filters. LTspice® simulation tool was used to design and simulate analog filters at 10 kHz. Maximum attenuation of-80 dB was achieved with the 7th order Chebyshev filter. As a result of the higher-order filter, the direct feedthrough effect was also minimized. The signal-To-noise ratio (SNR) would also be improved with a good receive section of the MPI scanner.

Original languageEnglish
Title of host publicationTIPTEKNO 2021 - Tip Teknolojileri Kongresi - 2021 Medical Technologies Congress
PublisherInstitute of Electrical and Electronics Engineers Inc.
ISBN (Electronic)9781665436632
DOIs
Publication statusPublished - 2021
Externally publishedYes
Event2021 Medical Technologies Congress, TIPTEKNO 2021 - Antalya, Turkey
Duration: 4 Nov 20216 Nov 2021

Publication series

NameTIPTEKNO 2021 - Tip Teknolojileri Kongresi - 2021 Medical Technologies Congress

Conference

Conference2021 Medical Technologies Congress, TIPTEKNO 2021
Country/TerritoryTurkey
CityAntalya
Period4/11/216/11/21

Bibliographical note

Publisher Copyright:
© 2021 IEEE.

Keywords

  • field of view
  • field-free point.
  • magnetic particle imaging