Skip to main navigation Skip to search Skip to main content

A tactile sensing capsule endoscope employing force sensing cantilevers for tumor diagnosis in the GI tract

  • Furkan Peker
  • , Gökalp Akcan*
  • , Dila Atak
  • , Ibrahim H. Ozata
  • , Derya S. Uymaz
  • , Emre Balik
  • , Müjdat Zeybel
  • , Hamdi Torun
  • , Onur Ferhanoğlu*
  • *Corresponding author for this work
  • Istanbul Technical University
  • Istanbul Bilgi University
  • Koc University
  • Northumbria University

Research output: Contribution to journal › Article › peer-review

3 Citations (Scopus)

Abstract

The correlation between tissue elasticity and histopathological diagnosis has brought attention to the development of biomedical devices for in-vivo measurement of tissue biomechanical properties. Towards this aim, we have developed a tactile sensing capsule endoscope to measure tissue Young’s modulus, in situ. Inspired by force microscopy, the capsule comprises four cantilevers that probe the walls of the GI tract using a single miniaturized actuator. The force exerted on the cantilever tip by the tissue is measured using the piezoelectric layer integrated on the cantilevers. The tactile-based modulus sensing capsule was initially tested on ex-vivo animal tissue, followed by healthy and cancerous human specimens. The results clearly delineate the differences in mechanical properties, with a Young’s modulus of 11.3 ± 2.3 kPa for healthy and 26.8 ± 4.6 kPa for cancerous tissue. Overall, in the realm of tactile-based modulus sensing of tissues, our technology uniquely combines localized, quantitative Young’s modulus measurements with the capability to perform multiple measurements throughout the GI tract wall in a single procedure. Moreover, the developed sensor has a compact form factor, in accordance with the capsule dimensions, and simple manufacturing steps using stereolithography. With further improvements, the developed medical device can be utilized as a non-invasive diagnostic tool in the clinic.

Original languageEnglish
Article number117391
JournalSensors and Actuators A: Physical
Volume399
DOIs
Publication statusPublished - 1 Mar 2026

Bibliographical note

Publisher Copyright:
© 2025 Elsevier B.V. All rights are reserved, including those for text and data mining, AI training, and similar technologies.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being
  2. SDG 9 - Industry, Innovation, and Infrastructure
    SDG 9 Industry, Innovation, and Infrastructure

Keywords

  • Biomechanical measurements
  • Capsule endoscopy
  • Early cancer diagnosis
  • Tactile sensing

Fingerprint

Dive into the research topics of 'A tactile sensing capsule endoscope employing force sensing cantilevers for tumor diagnosis in the GI tract'. Together they form a unique fingerprint.

Cite this