A low cost X-ray imaging device based on BPW-34 Si-PIN photodiode

E. Emirhan, A. Bayrak, E. Barlas Yücel, M. Yücel, C. S. Ozben*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

6 Citations (Scopus)

Abstract

A low cost X-ray imaging device based on BPW-34 silicon PIN photodiode was designed and produced. X-rays were produced from a CEI OX/70-P dental tube using a custom made ±30 kV power supply. A charge sensitive preamplifier and a shaping amplifier were built for the amplification of small signals produced by photons in the depletion layer of Si-PIN photodiode. A two dimensional position control unit was used for moving the detector in small steps to measure the intensity of X-rays absorbed in the object to be imaged. An Aessent AES220B FPGA module was used for transferring the image data to a computer via USB. Images of various samples were obtained with acceptable image quality despite of the low cost of the device.

Original languageEnglish
Pages (from-to)1-5
Number of pages5
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume819
DOIs
Publication statusPublished - 21 May 2016

Bibliographical note

Publisher Copyright:
© 2016 Elsevier B.V. All rights reserved.

Funding

We would like to thank to the Scientific and Technological Council of Turkey (TUBITAK) for supporting this work ( 113F104 ).

FundersFunder number
Scientific and Technological Council of Turkey
TUBITAK113F104

    Keywords

    • Low cost
    • Si-PIN photodiodes
    • X-ray detectors
    • X-ray imaging

    Fingerprint

    Dive into the research topics of 'A low cost X-ray imaging device based on BPW-34 Si-PIN photodiode'. Together they form a unique fingerprint.

    Cite this